Ambient Sensor

Water ORP Transmitter Detection Sensor

♦Adopt double high-impedance three-electrode system, with online one-key calibration, real-time temperature compensation, electrode loose alarm, electrode good or bad alarm during calibration, and pow

♦Adopt double high-impedance three-electrode system, with online one-key calibration, real-time temperature compensation, electrode loose alarm, electrode good or bad alarm during calibration, and power-off protection (the calibration result and preset data will not be lost due to shut down or power failure).

♦Over-current protection, over voltage protection, high measurement accuracy, fast response, long life and so on.

♦Small in size, light in weight, easy to install and maintain, and the standard industrial signal output (0-5V, 0-10V, 4-20MA, Modbus RTU485) can meet the limit of on-site connection to various real-time monitoring equipment.


Water ORP Transmitter Detection Sensor

ORP transmitter adopt double high-impedance three-electrode system, with online one-key calibration, real-time temperature compensation, electrode loose alarm, electrode good or bad alarm during calibration, and power-off protection (the calibration result and preset data will not be lost due to shut down or power failure) , Over-current protection, over voltage protection, high measurement accuracy, fast response, long life and so on.

This product is suitable for: electroplating, yeast, food processing, sewage treatment, metallurgy, environmental protection and other occasions.

ltemsParameters
Measurement ORP range-1000~1000mV or-2000~2000mV
Measuring ORP Accuracy±3mV@-1000~1000mV
±6mV@-2000~2000mV
ORP sensitivity1mV
Power supplyDC 12V-24V
Output0-5V/0-10V/4-20mA/RS485
load resistanceVoltage output: R load 2 10K Ω
Current output: R loads(Uvcc-310.02Q
Consumption<1W
Storage condition10-50℃(-20~+80°℃ peak) 20-60%RH
Working environment-20℃~+80℃
dimension65X46X28.5mm³

The linear relation between analog signals and the corresponding physical value:

0 to 5 v output:

(-1000~1000mV range):ORP=ORP output voltage*400-1000;

(-2000~2000mV range):ORP=ORP output voltage*800-2000;

0 to 10 v output:

(-1000~1000mV range):ORP=ORP output voltage*200-1000;

(-2000~2000mV range):ORP=ORP output voltage*400-2000;4-20 ma output:

(-1000~1000mV range):ORP=(ORP current-4)*125-1000;

(-2000~2000mV range):ORP=(ORP current-4)*250-1000;


Due to the ORP electrode in the measurement environment vulnerable to pollution cause electrode signal affected, in order to ensure the ORP sensor can normally use for a long time, usually need to make a correction on a regular basis through the calibration of transducer (once a month). Depending on the water samples of specific calibration cycle.

FAQ
Q1. Can I have a free sample for testing ?
A: Sample order is acceptable, but need you to pay for the product cost and freight .

Q2. What about the lead time?
A: Sample needs 4-5 working days after receiving your payment, mass production needs extra 4-5 working days.



Q3: Is it OK to print my logo on the products?
A: Yes, we support OEM/ODM service.


热门推荐